Hugues Lucron1,*, Alban-Elouen Baruteau2,3, Caroline Ovaert4, Ali Houeijeh5, Mélanie Brard1, Patrice Guerin2, François Bourlon6, Claire Dauphin7, Saskia Tuttle1, Maha Tagorti3, Rishika Banydeen8, François Godart5
Congenital Heart Disease, Vol.17, No.4, pp. 421-436, 2022, DOI:10.32604/chd.2022.020835
- 04 July 2022
Abstract Objectives: We aim to describe the efficacy, safety, and characteristics of the Amplatzer Vascular Plug (AVP) II and IV “off-label” use for multiple cardiovascular occlusions in children under 10 years. Methods: Observational retrospective multicenter (2007–2020, 6 centers) review of paediatric procedures using AVP II or IV. Results: A total of 125 children (49.6% aged ≤ 1 year, 147 lesions) underwent 136 successive procedures (success rate: 98.5%) using 169 devices (109 AVP IV, 60 AVP II). The mean device diameter was 7.7 ± 3.2 mm (4–20 mm). The median AVP size to vessel diameter ratio was 1.3 (0–2). The median… More >
Graphic Abstract