Jun Xie1, Chong Wang1, Jiaxiang Cai2, Fuhong Cai1,*
CMC-Computers, Materials & Continua, Vol.57, No.2, pp. 261-268, 2018, DOI:10.32604/cmc.2018.03669
Abstract The rapid quantification method of human serum glucose was established by using the Fourier transform infrared spectroscopy (FTIR) and attenuated total reflection (ATR). By the subtracted spectra between glucose aqueous solution and de-ionized water, absorption peaks are calculated in fingerprint area. Based on these absorption peaks and multiple linear regression (MLR) model, discrete band selection method of absorption peaks disturbance model (APDM) was developed. 5 absorption peaks 1150 cm-1, 1103 cm-1, 1078 cm-1, 1034 cm-1, 991 cm-1 were found in fingerprint area. Used these absorption peaks to establish absorption peaks disturbance model, the optimal wavelength combinations are 1140 More >