Table of Content

Open Access iconOpen Access

ABSTRACT

CFD case study to optimize surgical adjustment of ventricular assist device implantation to minimize stroke risk part I: steady-state CFD modeling

by Andres F. Osorio1, Ruben Osorio1, Reginald Tran1, Andres Ceballos1, Eduardo Divo2, I. Ricardo Argueta-Morales3

Department of Mechanical, Materials and Aerospace Engineering, University of Central Florida,Orlando, FL, USA.
Department of Engineering Technology, University of Central Florida, Orlando, FL, USA.
Congenital Heart Institute at Arnold Palmer Hospital, Orlando, FL, USA.
College of Medicine, University of Central Florida, Orlando, FL, USA.

The International Conference on Computational & Experimental Engineering and Sciences 2010, 15(4), 123-124. https://doi.org/10.3970/icces.2010.015.123

Abstract

Presently, mechanical support is the most promising alternative to cardiac transplantation. Ventricular Assist Devices (VADs) were originally used to provide mechanical circulatory support in patients waiting planned heart transplantation (“bridge-to-transplantation” therapy). The success of short-term bridge devices led to clinical trials evaluating the clinical suitability of long-term support (“destination” therapy) with left ventricular assist devices (LVADs). The first larger-scale, randomized trial that tested long-term support with a LVAD reported a 44% reduction in the risk of stroke or death in patients with a LVAD. In spite of the success of LVADs as bridge-to-transplantation and long-term support. Patients carrying these devices are still at risk of several adverse events. The most devastating complication is caused by embolization of thrombi formed within the LVAD or inside the heart into the brain. Despite anticoagulation management and improved LVAD design, there is still significant occurrence of thromboembolic events in patients. Investigators have reported that the incidence of thromboembolic cerebral events ranges from 14% to 47% over a period of 6-12 months.
Accepting the current rate of thrombus formation within the LVAD, an alternative method to reduce the incidence of cerebral embolization is hypothesized: thromboembolism to the carotid and vertebral arteries can be minimized by adjusting the placement of the LVAD outflow conduit, or by the placement of an aorticto-innominate artery bypass graft, or by the placement of an aortic-to-left-carotid artery bypass graft, or possibly a combination of these. We present a computational fluid dynamics (CFD) study of the aortic arch bed hemodynamics using a representative geometry of the human aortic arch and an alternative aortic bypass whose express purpose is to investigate the hypothesis.
We utilize the CFD code, STARCCM+, in which a Lagrangian particle-tracking model is coupled to the fluid flow solver to predict particle trajectories. In this study, we then explore the possibility of finding an optimal configuration of the LVAD conduit angle and anastomosis location (that is both distance from the innominate and polar angle in the coronal plane of the ascending aorta), and we seek to establish the benefits of the ligation of the innominate artery and placement of an aortic-to-innominate artery bypass graft or the ligation of the left carotid artery and placement of an aortic-to-left-carotid artery bypass graft. LVAD conduit flow is assumed to be steady (non-pulsatile), with negligible pulsatile flow originating from the aortic root. Although not always the case, steady or near steady flow characterizes the flow regime in many patient with LVAD support. Continuous-flow LVADs have predominantly become the device of choice. Consequently, the focus of this study is to investigate steady flow conditions.
Results are presented in the form of percentage of thrombi of diameters 2mm, 4mm and 5 mm reaching the carotid and vertebral arteries as a function of LVAD conduit placement and aortic bypass implantation, revealing promising improvement: significant (50%) reduction in thromboembolism by proper adjustment of the LVAD inflow cannula. We present adult and pediatric cases as well as a patient-specific model generated from image segmentation of a CT scan. We also discuss progress of a benchtop model devised to buttress the CFD study.

Cite This Article

APA Style
Osorio, A.F., Osorio, R., Tran, R., Ceballos, A., Kassab, A. et al. (2010). CFD case study to optimize surgical adjustment of ventricular assist device implantation to minimize stroke risk part I: steady-state CFD modeling. The International Conference on Computational & Experimental Engineering and Sciences, 15(4), 123-124. https://doi.org/10.3970/icces.2010.015.123
Vancouver Style
Osorio AF, Osorio R, Tran R, Ceballos A, Kassab A, Divo E, et al. CFD case study to optimize surgical adjustment of ventricular assist device implantation to minimize stroke risk part I: steady-state CFD modeling. Int Conf Comput Exp Eng Sciences . 2010;15(4):123-124 https://doi.org/10.3970/icces.2010.015.123
IEEE Style
A. F. Osorio et al., “CFD case study to optimize surgical adjustment of ventricular assist device implantation to minimize stroke risk part I: steady-state CFD modeling,” Int. Conf. Comput. Exp. Eng. Sciences , vol. 15, no. 4, pp. 123-124, 2010. https://doi.org/10.3970/icces.2010.015.123



cc Copyright © 2010 The Author(s). Published by Tech Science Press.
This work is licensed under a Creative Commons Attribution 4.0 International License , which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
  • 1183

    View

  • 770

    Download

  • 0

    Like

Share Link